품질 STORZ 13820pks 위장경 수리 공장

STORZ 13820pks 위장경 수리

STORZ 13820pks Gastroscope for Repair ► Brief Description Fault: 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of. Solution: 1. Replace the wires, coil pipe, rubber; 2.Adjust angle. ►Problem & Solutoin Brand STORZ Model 13820pks Fault 1.Rubber leakage; 2.Wire abrasion; 3. no up ...
품질 정비용 올림푸스 CF-H190L 콜로노스코프 공장

정비용 올림푸스 CF-H190L 콜로노스코프

Olympus CF-H190L Colonoscope for repair ► Brief Description Fault: 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; ...
품질 삼성 메디슨 P2-4AH 심혈관 검출기 수리 공장

삼성 메디슨 P2-4AH 심혈관 검출기 수리

Samsung Medison P2-4AH Cardiac probe for repair ► Brief Description Fault:Lens problem, crystal damage. Solution:Replace probe lens, replace crystal. ►Problem & Solutoin Brand Samsung Model P2-4AH Fault Lens problem, crystal damage Solution Replace probe lens, replace crystal ► Common issues with P2...
품질 정비용 OLYMPUS TJF-Q180V DUODENOSCOPE 공장

정비용 OLYMPUS TJF-Q180V DUODENOSCOPE

OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR ► Brief Description Fault: 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. Solution: Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and ...
품질 고쳐야 할 올림푸스 CF-240L 콜로노스코프 공장

고쳐야 할 올림푸스 CF-240L 콜로노스코프

Olympus CF-240L Colonoscope for repair ► Brief Description Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch. Solution: Replace rubber, objective lens, C cover. Adjust the angle. ►Problem & Solutoin Brand Olympus Model CF-240L Fault Leakage...
품질 카를 스토르츠 27005BA 내시경 수리 공장

카를 스토르츠 27005BA 내시경 수리

Karl Storz 27005BA endoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
품질 도시바 PST-25BT 위상 배열 프로브 수리용 공장

도시바 PST-25BT 위상 배열 프로브 수리용

Toshiba PST-25BT Phased Array Probe For Repair ► Brief Description Fault: Housing break, image weak, crystal aging. Solution: Replace crystal and housing. ►Problem & Solutoin Brand Toshiba Model PST-25BT Fault Housing break, image weak, crystal aging Solution Replace crystal and housing ► Common ...
품질 STRYKER 1188-210-105 카메라 헤드 수리용 공장

STRYKER 1188-210-105 카메라 헤드 수리용

STRYKER 1188-210-105 camera head for repair ► Brief Description Fault: Image with interrupt. Cable aging. Solution: Replace compatible new cable and connector. ►Problem & Solutoin Brand STRYKER Model 1188-210-105 Fault Image with interrupt. Cable aging Solution Replace compatible new cable and ...
품질 STORZ 카메라 헤드 H3-ZA 수리 공장

STORZ 카메라 헤드 H3-ZA 수리

STORZ Camera Head H3-ZA For Repair ► Brief Description Fault:Image streaks frequently,cable loose contact and aged. Solution: Replace cable kit. ►Problem & Solutoin Brand STORZ Model H3-ZA Fault Image streaks frequently,cable loose contact and aged Solution Replace cable kit ► Common issues with H3...
품질 정비용 GE RIC6-12-D 내구성 탐사선 공장

정비용 GE RIC6-12-D 내구성 탐사선

GE RIC6-12-D endocavity probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens,Clean cable. ►Problem & Solutoin Brand GE Model RIC6-12-D Fault Lens problem Solution Replace lens,Clean cable ► Common issues with RIC6-12-D Probe Dome Damage and Fluid Leaks Image Dropouts ...
품질 히타치 C41V1 초음파 탐사선 수리 공장

히타치 C41V1 초음파 탐사선 수리

Hitachi C41V1 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Brand Hitachi Model C41V1 Fault Lens problem Solution Replace lens ► Common issues with C41V1 Probe Image Dropouts or Streaks Cable and Connector Damage Acoustic Lens ...
품질 보정용 소노스케이프 8-L743 선형 프로브 공장

보정용 소노스케이프 8-L743 선형 프로브

Sonoscape 8-L743 linear probe for repair ► Brief Description Fault: Crystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model 8-L743 Fault Crystal problem Solution Replace the crystal ► Common issues with 8-L743 Probe Acoustic Lens Damage Piezoelectric Crystal Damage ...
품질 소노스케이프 C362 복부 론드 수리 공장

소노스케이프 C362 복부 론드 수리

Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens ...
품질 Careutech BK Medical 9048 초음파 프로브 수리용 공장

Careutech BK Medical 9048 초음파 프로브 수리용

BK Medical 9048 Ultrasound Probe For Repair ► Brief Description Fault: The housing break. Solution: Replace probe housings. ►Problem & Solutoin Brand BK Medical Model 9048 Fault The housing break. Solution Replace probe housings. ► Common issues with 9048 Probes No Image / Poor Image Quality Probe ...
품질 의료 장비 부품 소노스케이프 L743 수리용 선형 탐사 공장

의료 장비 부품 소노스케이프 L743 수리용 선형 탐사

Sonoscape L743 linear probe for repair ► Brief Description Fault: Probe head broken. Solution: Replace probe head.(including crystal and upper housing). ►Problem & Solutoin Brand Sonoscape Model L743 Fault Probe head broken Solution Replace probe head.(including crystal and upper housing) ► Common ...
품질 BK 메디컬 9096 로봇 초음파 프로브 수리용 공장

BK 메디컬 9096 로봇 초음파 프로브 수리용

BK Medical 9096 Robotic Ultrasound Probe For Repair ► Brief Description Fault: All the outer skin of the cable is seriously damaged. Solution: Replace the outer skin of the cable. ►Problem & Solutoin Brand BK Medical Model 9096 Fault All the outer skin of the cable is seriously damaged Solution ...
품질 삼성 메디슨 3D4-9 프로브 SN P0N7 수리 공장

삼성 메디슨 3D4-9 프로브 SN P0N7 수리

Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D...
품질 의료기기 수리 GE RIC5-9-D 초음파 검사의 수리용 케이스 공장

의료기기 수리 GE RIC5-9-D 초음파 검사의 수리용 케이스

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...
품질 앵글 와이어 및 헬릭스 튜브 착용 CF-HQ190L 20% 깨진 라이트 펀들 공장

앵글 와이어 및 헬릭스 튜브 착용 CF-HQ190L 20% 깨진 라이트 펀들

Products: CF-HQ190L For Repair Problem & Solutoin: CF-HQ190L - Light bundle 20% Broken Knob Leaked The angle wire and helix tube Worn Biospy channel Worn Rubber Cracked Light guide tube Indented Solution Replace Biopsy Channel, Wire, Light Guide Tube, Knob, Rubber Applications: When it comes to ...
품질 의료기기 수리 케이스 오림푸스 CF-H190L 수리용 콜로노스코프 공장

의료기기 수리 케이스 오림푸스 CF-H190L 수리용 콜로노스코프

Olympus CF-H190L Colonoscope for repair,SN:261****. Fault:Light bundle broken,biopsy tube worn,water air tube leakage,bending section deformation,wire helix tube worn. Solution:Replace the bending section,wire,helix tube,air tube and rubber.PS:Warranty for replace parts only. Repair time:7 working ...
품질 수리된 올림푸스 의료 장비 의료 서비스의 궁극적인 해결책 공장

수리된 올림푸스 의료 장비 의료 서비스의 궁극적인 해결책

Product Description: Product Summary: The Medical Device Repair Cases product offered is a comprehensive solution for repairing and maintaining Olympus medical devices. With a warranty of 90 days, customers can have peace of mind knowing that their repaired devices are covered for an extended period...
품질 OLYMPUS CF-HQ190L 의료기기 수리 케이스 수리 도구 케이스 공장

OLYMPUS CF-HQ190L 의료기기 수리 케이스 수리 도구 케이스

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...