| 상표 | 필립스 |
|---|---|
| 모델 | L15-7io |
| 잘못 | 렌즈 문제 |
| 해결책 | 프로브 렌즈를 교체하십시오 |
| 포장 세부 사항 | 표준 패키지 |
| Brand | OLYMPUS |
|---|---|
| Model | TJF-Q180V |
| Fault | 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. |
| Solution | Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and rubber. (Nowarranty for CCD) |
| 포장 세부 사항 | 표준 패키지 |
| Brand | Olympus |
|---|---|
| Model | CF-240L |
| Fault | Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch, |
| Solution | Replace rubber, objective lens, C cover. Adjust the angle. |
| Packaging Details | Standard Package |
| Brand | Karl Storz |
|---|---|
| Model | 27005BA |
| Fault | 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. |
| Solution | 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope a |
| Packaging Details | Standard Package |
| 브랜드 | 토시바 |
|---|---|
| 모델 | PST-25BT |
| 결점 | 주택 파손, 이미지 약화, 결정 노후화 |
| 해결책 | 크리스탈과 하우징을 교체하십시오 |
| 포장 세부 사항 | 표준 패키지 |
| 브랜드 | 스트라이커 |
|---|---|
| 모델 | 1188-210-105 |
| 결점 | 인터럽트가있는 이미지. 케이블 노화 |
| 해결책 | 호환 새 케이블 및 커넥터를 교체하십시오 |
| 포장 세부 사항 | 표준 패키지 |
| 브랜드 | 스토르츠 |
|---|---|
| 모델 | H3-ZA |
| 결점 | 이미지 줄무늬가 자주, 케이블 느슨한 접촉 및 숙성 |
| 해결책 | 케이블 키트를 교체하십시오 |
| 포장 세부 사항 | 표준 패키지 |
| 상표 | GE |
|---|---|
| 모델 | RIC6-12-D |
| 잘못 | 렌즈 문제 |
| 해결책 | 렌즈, 깨끗한 케이블을 교체하십시오 |
| 포장 세부 사항 | 표준 패키지 |
| 상표 | 메드트로닉 |
|---|---|
| 모델 | 1898200T |
| 잘못 | 1) 베어링이 부식되고 밀폐가 깨졌습니다. 2) 엔진이 고정되어, 내부는 살균 잔해, 모터 부식으로 가득 차 있습니다. 3) 칼 머리의 회전 방향이 고정되어 있습니다.4) 페인트 |
| 해결책 | 케이블을 제외한 핸드 피스 전체를 교환하십시오. |
| 포장 세부 사항 | 표준 패키지 |